首页> 外文会议>International Congress on Image and Signal Processing >Evaluation of Modulation Transfer Function on the optical tomography imaging system
【24h】

Evaluation of Modulation Transfer Function on the optical tomography imaging system

机译:在光学层析成像系统上评估调制传递函数

获取原文

摘要

The optical tomography system based on the single pixel detector produces the high resolution image with optical scanning. This paper introduces a method to evaluate the quality of the optical tomography system with MTF (Modulation Transfer Function). At first use the calibration board that are digital images for software simulation, and then calculate the MTF values on the different space frequency in order to analyze the change of MTF and the reasons of image degradation. Followed by using the integrating sphere light source and film calibration boards in experiments, it tests the quality of the tomography imaging system and the validation of imaging properties. The system produces the images with Filter Back Projection (FBP). It analyzes image diffusion situation when the target image is the round hole. The results suggest that the effect of algorithm on the reconstructed image for the optical tomography system cannot be ignored.
机译:基于单像素检测器的光学层析成像系统可通过光学扫描生成高分辨率图像。本文介绍了一种使用MTF(调制传递函数)评估光学层析成像系统质量的方法。首先使用作为数字图像的校准板进行软件仿真,然后在不同的空间频率上计算MTF值,以分析MTF的变化和图像退化的原因。随后在实验中使用积分球光源和胶片校准板,它测试了层析成像系统的质量和成像特性的验证。系统使用反投影(FBP)生成图像。分析目标图像为圆孔时的图像扩散情况。结果表明,算法对光学层析成像系统重建图像的影响不容忽视。