首页> 外文会议>International Conference on BioMedical Engineering and Informatics >Improved positioning methods based on MR images in the MR-guided PHIFU system
【24h】

Improved positioning methods based on MR images in the MR-guided PHIFU system

机译:MR引导PHIFU系统中基于MR图像的改进定位方法

获取原文

摘要

The positioning of the ultrasound system is the key problem when using the MR to guide the pHIFU(phased High Intensity Focused Ultrasound) system. This paper will discuss two methods to improve the transfer relationship between the ultrasound coordinate and the MR coordinate. First, if taking the probe as a media to guide the positioning, additional work need to reduce calculating errors including improved designs of the markers on the probe and the design of redundancy of markers. Second, if taking the thermometric data from MR as a media to guide the positioning, another transfer matrix can be built between the ultrasound coordinate and the MR coordinate to calibrate outcomes of the primary location. In terms of experiment results, it's proved that full use of redundant information from more than four markers can improve the accuracy of positioning obviously. And we will do further research to test the outcomes of another method.
机译:超声系统的定位是使用MR引导pHIFU(阶段性高强度聚焦超声)系统时的关键问题。本文将讨论两种改善超声坐标和MR坐标之间传递关系的方法。首先,如果以探针作为引导定位的媒介,则需要减少额外的工作来减少计算误差,包括改进探针上标记的设计以及标记的冗余设计。其次,如果将来自MR的测温数据作为引导定位的介质,则可以在超声坐标和MR坐标之间建立另一个传递矩阵,以校准主要位置的结果。实验结果表明,充分利用四个以上标记的冗余信息可以明显提高定位精度。我们将做进一步的研究,以测试另一种方法的结果。