首页> 外文会议>International Conference on Electronics Information and Emergency Communication >Analysis of non-insertable device under test measurement error with vector network analyzer
【24h】

Analysis of non-insertable device under test measurement error with vector network analyzer

机译:用矢量网络分析仪分析非插入式被测设备的测量误差

获取原文

摘要

The paper introduces the 12 error terms calibration model algorithm of vector network analyzer(VNA) based on the signal flow graph theory and the Mason formula, and proposes the full 2-port calibration method. The influence of the error caused by measuring non-insertable DUTs with the VNA to the measuring results is analyzed and a corresponding compensation algorithm is given. Meanwhile, this paper discusses how the transmission measurement error is affected respectively by hardware performance of the VNA, the reflection coefficient and transmission coefficient of the through calibration kit and the S parameters of the DUT. Finally, an algorithm which can eliminate the unknown through is presented.
机译:介绍了基于信号流图理论和梅森公式的矢量网络分析仪(VNA)的十二项误差项标定模型算法,并提出了全二端口标定方法。分析了用VNA测量不可插入DUT所引起的误差对测量结果的影响,并给出了相应的补偿算法。同时,本文讨论了传输测量误差如何分别受到VNA的硬件性能,直通校准套件的反射系数和传输系数以及DUT的S参数的影响。最后,提出了一种可以消除未知现象的算法。