A simplified method for quantifying the subject-specific relationship between blood pressure and carotid-femoral pulse wave velocity

机译:一种简化的方法,用于量化血压与颈动脉脉搏波速度之间的主题特异性关系

获取原文

摘要

Devices that estimate blood pressure from arterial pulse wave velocity (PWV) potentially provide continuous, ambulatory blood pressure monitoring. Accurate blood pressure estimation requires reliable quantification of the relationship between blood pressure and PWV. Regression to population normal values or, when using limb artery PWV, changing hydrostatic blood pressure within the limb provides a calibration index. Population lookup tables require accurate anthropometric correlates, assuming no individual variation. Only devices that measure PWV in the limb can use limb position changes. This study proposes a method for developing a calibration curve independent of lookup tables and useful for large artery PWV measurement, such as carotid-femoral PWV (PWV). PWV was measured in 27 normal subjects (15 female, 36±19 years) in both the supine and standing position. The change in systemic pressure was measured and hydrostatic pressure change calculated from estimated vessel path length height, measured using body surface distances. Brachial diastolic blood pressure increased for all subjects from supine to standing (supine 70±8 mmHg, standing 83±8 mmHg, p<;0.001) with an additional hydrostatic change across the carotid-femoral path length of 19±2 mmHg (p<;0.001). PWV also increased in all subjects (supine 5.2±1.3 m/s, standing 7.3±2.2 m/s, p<;0.001). The subject-specific calibration index (ΔDP/ΔPWV) varied amongst the cohort (20±8 mmHg/m/s), was correlated with age (-0.57, p=0.002) and seated aortic systolic pressure (-0.38, p=0.048) and was always greater than zero. Thus, this study describes a simple but novel method of measuring an individualized calibration index using blood pressure and PWV measurements in the supine and standing position.
机译:从动脉脉冲波速度(PWV)估计血压的装置可能提供连续,动态血压监测。精确的血压估计需要可靠地量化血压和PWV之间的关系。对群体正常值的回归或使用肢体动脉PWV时,在肢体内改变静液压提供校准指数。人口查找表需要准确的人体计量相关,假设没有单独的变化。只有测量肢体中PWV的设备可以使用肢体位置变化。该研究提出了一种用于开发独立于查找表的校准曲线的方法,可用于大型动脉PWV测量,例如颈动脉股PWV(PWV)。在仰卧和立场的27个正常受试者(15名女性,36±19年)中测量PWV。测量系统压力的变化,并根据估计的血管路径长度高度计算的静压压力变化,使用体表距离测量。从仰卧到静止的所有受试者增加(仰卧70±8 mmHg,静置83±8 mmHg,p <; 0.001),颈部股径长度为19±2 mmHg(P <) ; 0.001)。所有受试者的PWV也增加(仰卧5.2±1.3 m / s,静置7.3±2.2 m / s,p <; 0.001)。在队列(20±8mmHg / m / s)中变化的主题特异性校准指数(Δdp/Δpwv)与年龄(-0.57,p = 0.002)和坐的主动脉收缩压(-0.38,p = 0.048 )并且总是大于零。因此,本研究描述了一种使用血压和仰卧位的PWV测量来测量个性化校准指数的简单但新的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号