首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Design, development and experimental validation of a non-invasive device for recording respiratory events during bottle feeding
【24h】

Design, development and experimental validation of a non-invasive device for recording respiratory events during bottle feeding

机译:用于记录奶瓶喂养过程中呼吸事件的无创设备的设计,开发和实验验证

获取原文

摘要

In newborns, a poor coordination between sucking, swallowing and breathing may undermine the effectiveness of oral feeding and signal immaturity of Central Nervous System. The aim of this work is to develop and validate a non-invasive device for recording respiratory events of newborns during bottle feeding. The proposed device working principle is based on the convective heat exchanged between two hot bodies and the infants' breathing. The sensing elements are inserted into a duct and the gas exchanged by infants is conveyed into this duct thanks to an ad hoc designed system to be mounted on a commercial feeding bottle. Two sets of experiments have been carried out in order to investigate the discrimination threshold of the device and characterize the sensor response at oscillating flows. The effect of distance and tilt between nostrils and device, and the breathing frequency, have been investigated simulating nostrils and neonatal respiratory pattern. The device has a discrimination threshold lower than 0.5 L/min at both 10° and 20° of tilt. Distance for these two settings does not affect the threshold in the investigated range (10–20 mm). Moreover, the device is able to detect breathing events, and to discriminate the onset of expiratory phase, during a neonatal respiratory task delivered by a lung simulator. The results foster the successful application of this device to the assessment of the temporal breathing pattern of newborns during bottle feeding with a non-invasive approach.
机译:在新生儿中,吮吸,吞咽和呼吸之间的协调不良可能会破坏口服喂食的有效性,并削弱中枢神经系统的信号。这项工作的目的是开发和验证一种无创设备,用于记录奶瓶喂养期间新生儿的呼吸事件。所提出的装置工作原理是基于两个热体之间的对流热交换和婴儿的呼吸。感测元件插入管道中,婴幼儿交换的气体通过专门设计的系统安装在商用奶瓶上,被输送到该管道中。为了调查设备的判别阈值并表征振荡流量下的传感器响应,已进行了两组实验。已经研究了鼻孔和装置之间的距离和倾斜度以及呼吸频率的影响,以模拟鼻孔和新生儿的呼吸模式。该器件在10°和20°倾斜时的辨别阈值均低于0.5 L / min。这两个设置的距离不会影响所研究范围内的阈值(10–20 mm)。此外,在由肺模拟器进行的新生儿呼吸任务期间,该装置能够检测呼吸事件并区分呼气阶段的开始。结果促进了该设备在无奶喂养奶瓶喂养过程中成功地用于评估新生儿的短暂呼吸模式。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号