首页> 外文会议>International Conference on BioMedical Engineering and Informatics >Ophthalmological ultrasound biometer using Golay-coded pulse excitation
【24h】

Ophthalmological ultrasound biometer using Golay-coded pulse excitation

机译:使用格雷编码脉冲激励的眼科超声生物测定仪

获取原文

摘要

The requirements of acoustic output parameters in clinical instruments are very strict. This paper proposed a method that introducing the Golay complementary coded excitation and decoded compression to ophthalmological ultrasound biometer, instead of the traditional one using one pulse excitation. By adopting this method, the excitation voltage can be cut down from 60 V to ±15 V. Thus, the ultrasonic instantaneous excitation power was reduced greatly, and in the meantime the average excitation power was enhanced. A one reflection experiment, a resolution measurement experiment and in vivo experiment were designed in this paper. The results showed that when cutting down the excitation voltage but using Golay-coded pulse excitation can also maintain the same detective depth, signal intensity, resolution and measuring accuracy as the high excitation voltage.
机译:临床仪器中对声音输出参数的要求非常严格。本文提出了一种将Golay互补编码激励和解码压缩引入眼科超声生物计的方法,代替了传统的使用单脉冲激励的方法。通过采用这种方法,可以将激励电压从60 V降低到±15V。因此,超声瞬时激励功率大大降低了,同时平均激励功率也得到了提高。本文设计了一个单反射实验,一个分辨率测量实验和一个体内实验。结果表明,在降低激励电压时,使用Golay编码的脉冲激励还可以保持与高激励电压相同的检测深度,信号强度,分辨率和测量精度。