首页> 外文会议>IEEE International Conference on Nano/Micro Engineered and Molecular Systems >Cantilever arrayed blood pressure sensor for arterial applanation tonometry
【24h】

Cantilever arrayed blood pressure sensor for arterial applanation tonometry

机译:悬臂阵列血压传感器用于动脉压平眼压计

获取原文

摘要

We developed a cantilever-arrayed blood pressure sensor array fabricated by (111) silicon bulk-micromachining for the noninvasive and continuous measurement of blood pressure. The blood pressure sensor measures the blood pressure based on the change in resistance of the piezoresistor on a 5-µm-thick-arrayed perforated membrane and 20-µm-thick metal pads. The length and width of the unit membrane are 210 and 310 µm, respectively. The width of the insensible zone between adjacent units is only 10 µm. The resistance change over contact force was measured to verify the performance. The good linearity of the result confirmed that the PDMS package transfers the forces appropriately. The measured sensitivity was about 4.5%/N. The maximum measurement range and resolution of the fabricated blood pressure sensor were greater than 900 mmHg and less than 1 mmHg, respectively.
机译:我们开发了一种由(111)硅块微机械加工制成的悬臂阵列式血压传感器阵列,用于无创且连续地测量血压。血压传感器根据在厚度为5 µm的多孔膜和厚度为20 µm的金属垫上的压敏电阻的电阻变化来测量血压。单位膜的长度和宽度分别为210和310 µm。相邻单元之间的感官区域的宽度仅为10 µm。测量电阻在接触力上的变化以验证性能。结果的良好线性证明了PDMS封装可以适当地传递力。测得的灵敏度约为4.5%/ N。所制造的血压传感器的最大测量范围和分辨率分别大于900 mmHg和小于1 mmHg。