The purpose of this study was to assess two methods of calibration of Respiratory Inductive Plethysmography (RIP), isovolume (ISOCAL) and qualitative diagnostic calibration (QDC), in two-body positions (sitting and standing). Twenty-three normal subjects (15 females, 8 males) aged 25,4 ± 3,9 (mean ± standard deviation) years were monitored. The results indicated that the calibration factor (K) is significantly different between both calibration methods. The K mean values, and standard deviation, corresponding to ISOCAL and QDC were 2,6 ± 0,8 and 0,7 ± 0,2 for standing position, and 2,0 ± 0,8 and 0,6 ± 0,2 for sitting position, respectively. We obtained K values in ISOCAL calibration compatible to physiological understanding of the body positions applied.
机译:这项研究的目的是评估在两个身体位置(坐姿和站立姿)的两种方法:等体积(ISOCAL)和定性诊断校准(QDC)。监测了二十三名正常受试者(15名女性,8名男性),年龄25.4±3,9(平均±标准差)岁。结果表明,两种校准方法之间的校准因子(K)显着不同。对应于ISOCAL和QDC的K平均值和标准偏差对于站立位置分别为2.6±0.8和0.7±0.2,以及2.0±0.8和0.6±0.2分别用于坐姿。我们在ISOCAL校准中获得了K值,可与对所应用身体位置的生理理解相兼容。