Radiation dose measurement on computed tomography scanner using thermoluminescent dosemeter

机译:使用热莫荧光剂量计对计算机断层扫描仪的辐射剂量测量

获取原文

摘要

Computed Tomography scan is a method applied to examine the anatomy of a human body. A CT scan is also used as an alternative way to detect cancer. However due to the utilization of X-Rays, organs being examined are considered to be at risk of radiation effect. The purpose of this research is to measure absorbed radiation dose to meet the requirements of radiation protection on patients. A solid humanoid phantom was used to simulate the patient body. For the dose measurement, a lithium fluoride thermoluminescent dosemeters (TLD-100) were mounted within the head and neck area of the phantom. This research is also aimed to make the commonly used scanning protocol within a hospital to its full potential. This is done by comparing the radiation dose between different scanning protocols to find the smallest radiation exposure. It can be concluded that, the thicker the slice, the lower the exposed radiation dose will be.
机译:计算机断层扫描扫描是应用用于检查人体解剖结构的方法。 CT扫描也被用作检测癌症的替代方法。然而,由于利用X射线,被检查的器官被认为存在辐射效应的风险。本研究的目的是测量吸收的辐射剂量,以满足患者辐射保护的要求。用于模拟患者体的固体的人形幽灵。对于剂量测量,氟化锂热敏发光剂仪(TLD-100)安装在幻像的头部和颈部区域内。该研究还旨在使医院内的常用扫描协议成为其全部潜力。这是通过比较不同扫描协议之间的辐射剂量来找到最小的辐射曝光来完成的。可以得出结论,切片越厚,暴露的辐射剂量越低。