首页> 外文会议>33rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Assessment of the effect of time in the repeatability of the stabilometric parameters in diabetic and non-diabetic subjects during bipedal standing using the LorAn pressure distribution measurement system
【24h】

Assessment of the effect of time in the repeatability of the stabilometric parameters in diabetic and non-diabetic subjects during bipedal standing using the LorAn pressure distribution measurement system

机译:使用LorAn压力分布测量系统评估时间对双足站立时糖尿病和非糖尿病受试者稳定参数重复性的影响

获取原文

摘要

This study was designed to assess the effect of time on the repeatability of the LorAn pressure distribution measurement system, and evaluate the variability of plantar pressure and postural balance, during barefoot standing in diabetic and non-diabetic subjects, for future diabetic foot clinical evaluation. Fourteen subjects were evaluated (8 females, 6 males, 8 non-diabetics and 6 diabetics, age range 30–70 years) and had no musculoskeletal symptoms. Four variables were measured with the platform in the barefoot standing position. Ten measurements were taken using two different techniques for feet and posture positioning, during three sessions, once a week. The MANOVA test confirmed that the platform measurements are reproducible for variables body baricenter (x) and foot baricenter (x) through time, being the coefficients of variation, with a 99% confidence interval, lower than 1.6% for body baricenter (x), and lower than 2.06% for foot baricenter (x), for all studied conditions. For the remaining variables, the results were not stabilized through time, which makes necessary to standardize the measurement protocol that guarantees the repeatability in all variables.
机译:本研究旨在评估时间对LorAn压力分布测量系统的可重复性的影响,并评估在糖尿病和非糖尿病患者的赤脚站立过程中足底压力和姿势平衡的变异性,以用于将来的糖尿病足临床评估。评估了14位受试者(8位女性,6位男性,8位非糖尿病患者和6位糖尿病患者,年龄范围30-70岁),并且没有肌肉骨骼症状。在平台赤脚站立的状态下测量了四个变量。每周一次,在三个阶段中,使用两种不同的技术对脚和姿势进行了十次测量。 MANOVA测试证实,平台测量对于随时间变化的身体重心(x)和脚重心(x)是可重复的,是变异系数,置信区间为99%,低于身体重心(x)的1.6%,在所有研究条件下,脚重心(x)均低于2.06%。对于其余变量,结果随着时间的推移而不稳定,因此有必要对保证所有变量可重复性的测量协议进行标准化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号