首页> 外文会议>2011 4th International Conference on Biomedical Engineering and Informatics >Validation of an automated dynamic phantom to assess the performance of ultrasound system
【24h】

Validation of an automated dynamic phantom to assess the performance of ultrasound system

机译:验证自动动态体模以评估超声系统的性能

获取原文

摘要

There are many different approaches to evaluate image quality of ultrasound systems. Usually these tests are carried out manually, they are very time consuming, the results are subjective and repeatability can be poor. As one proposed improvement, the resolution integral produces one single figure of merit for image quality. The work reported here is based on a novel approach for grey-scale testing, based on the resolution integral, which can produce non-subjective and repeatable results for each combination of imaging system and transducer array. A total of 21 systems have been tested with three different depth settings to demonstrate its capabilities.
机译:有许多不同的方法可以评估超声系统的图像质量。通常,这些测试是手动执行的,非常耗时,结果主观,可重复性很差。作为一项提议的改进,分辨率积分可为图像质量产生一个单一的品质因数。此处报告的工作基于基于分辨率积分的灰度测试的新方法,该方法可为成像系统和换能器阵列的每种组合产生非主观的和可重复的结果。总共对21个系统进行了三种不同深度设置的测试,以证明其功能。