首页> 外文会议>2011 1st Middle East Conference on Biomedical Engineering >Detection and processing of full channel ionic current through indigenized Patch Clamp Technique
【24h】

Detection and processing of full channel ionic current through indigenized Patch Clamp Technique

机译:本地化膜片钳技术检测和处理全通道离子电流

获取原文
获取外文期刊封面目录资料

摘要

The importance of Noble prize winning “Patch Clamp Technique” is well documented. However, Patch Clamp Technique is very expensive and hence hinders research in developing countries. In this paper detection, processing and recording of ultra low current from induced cells by using transimpedence amplifier is described. The sensitivity of the proposed amplifier is in the range of femto amperes (fA). Capacitive-feedback is used with active load to obtain a 20MΩ transimpedance gain. The challenging task in designing includes achieving adequate performance in gain, noise immunity and stability. The circuit designed by the authors was able to measure current in the range of 300 fA to 100 pA. Adequate performance shown by the amplifier with different input current and outcome result was found to be within the acceptable error range. Results were recorded using LabView 8.5® for further research.
机译:贵族屡获殊荣的“膜片钳技术”的重要性已得到充分证明。但是,膜片钳技术非常昂贵,因此阻碍了发展中国家的研究。在本文中,描述了使用跨阻放大器检测,处理和记录来自感应细胞的超低电流。拟议的放大器的灵敏度在毫微安培(fA)的范围内。电容反馈与有功负载配合使用可获得20MΩ的跨阻增益。设计中具有挑战性的任务包括在增益,抗噪性和稳定性方面获得足够的性能。作者设计的电路能够测量300 fA至100 pA的电流。放大器在不同的输入电流和结果下显示出的足够性能被认为在可接受的误差范围内。使用LabView 8.5 ®记录结果以供进一步研究。