首页> 外文会议>32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Preliminary assessment of abdominal organ perfusion utilizing a fiber optic photoplethysmographic sensor
【24h】

Preliminary assessment of abdominal organ perfusion utilizing a fiber optic photoplethysmographic sensor

机译:利用光纤光电容积描记器传感器初步评估腹部器官灌注

获取原文

摘要

In an attempt to overcome the limitations of current techniques for monitoring abdominal organ perfusion, a prototype reflectance fiber optic photoplethysmographic (PPG) sensor and processing system was evaluated on seventeen anaesthetized patients undergoing laparotomy. Good quality PPG signals were obtained from the large bowel, small bowel, liver and stomach. Simultaneous PPG signals from the finger were also obtained for comparison purposes using an identical fiber optic sensor. Analysis of the mean ac and dc PPG amplitudes of all acquired signals indicated larger amplitudes for those signals obtained from abdominal organs than those obtained from the finger. Mean estimated blood oxygen saturation (SpO2) values from all abdominal sites showed good agreement with those obtained from the finger using both the finger fiber optic sensor and a commercial finger pulse oximeter. Furthermore, a Bland and Altman between-method-differences analysis on the estimated SpO2 data suggests that a fiber optic abdominal sensor may be a suitable method for the evaluation of abdominal organ perfusion.
机译:为了克服当前监测腹部器官灌注的技术的局限性,对十七名接受剖腹手术的麻醉患者进行了原型反射光纤光体积描记器(PPG)传感器和处理系统的评估。从大肠,小肠,肝脏和胃中获得了高质量的PPG信号。为了进行比较,还使用相同的光纤传感器从手指获得了同时的PPG信号。对所有采集信号的平均交流和直流PPG幅度的分析表明,从腹部器官获得的那些信号的幅度要比从手指获得的幅度更大。所有腹部部位的平均估计血氧饱和度(SpO 2 )值与使用手指光纤传感器和商用手指脉搏血氧仪从手指获得的血氧饱和度值显示出良好的一致性。此外,对估计的SpO 2 数据进行的Bland和Altman方法差异分析表明,光纤腹部传感器可能是评估腹部器官灌注的合适方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号