首页> 外文会议>Human Factors and Ergonomics Society annual meeting;HFES 2008 >Improving Neutron Detection: Usability Analysis of Control Mechanisms and Control Orientations for use in Hand-Held Devices
【24h】

Improving Neutron Detection: Usability Analysis of Control Mechanisms and Control Orientations for use in Hand-Held Devices

机译:改善中子探测:用于手持设备的控制机构和控制方向的可用性分析

获取原文

摘要

This study was conducted to determine the best control mechanism for a hand-held neutron detector testing three control mechanisms (blister buttons, push-buttons and a single-axis rocker) and two control mechanism orientations (vertical and horizontal). A simple menu-selection task, based on the current neutron detector, was conducted to evaluate the prototypes and determine the most efficient combination of control mechanisms and handle shapes. The most efficient combination was based on the lowest error rate and movement time. The results of this investigation found support for the horizontally oriented controls. In addition, blister buttons were found to be the best control mechanism (lowest error) out of the three control mechanisms tested. Blister buttons should have a diameter of at least 1.3 cm and provide tactile feedback to the first responder in order to enhance user performance.
机译:进行这项研究是为了确定手持中子探测器测试三种控制机构(气泡按钮,按钮和单轴摇杆)和两种控制机构方向(垂直和水平)的最佳控制机制。进行了基于当前中子探测器的简单菜单选择任务,以评估原型并确定控制机制和手柄形状的最有效组合。最有效的组合是基于最低的错误率和移动时间。这项调查的结果为水平方向的控件提供了支持。此外,在测试的三种控制机制中,发现气泡按钮是最好的控制机制(最低错误)。吸塑按钮的直径至少应为1.3厘米,并向第一响应者提供触觉反馈,以增强用户性能。

著录项