This paper describes the amendment to ISO 16063-22 that allows secondary calibrations of accelerometers up to 200,000 g_n, the current maximum range for commercial accelerometers. The amendment increases secondary calibration amplitudes for Hopkinson bar testing from 10,000 g_n to 200,000 g_n. These calibrations may be performed by any accelerometer manufacturer or any laboratory that has responsibility for accelerometer calibration. Additionally, the amendment provides guidance for the minimum pulse duration to avoid dispersion. This amendment is approved as of June 2012 and will be considered as a final revision during the upcoming ISO TC108 Meeting scheduled in Seoul, South Korea, 2013.
机译:本文介绍了ISO 16063-22的修正,允许加速度计的二次校准高达200,000 G_N,商业加速度计的电流最大范围。 该修正案增加了霍普金森棒的二级校准幅度从10,000 G_N到200,000 G_N测试。 这些校准可以由任何加速度计制造商或任何负责加速度计校准的实验室进行的。 此外,该修正案为最低脉冲持续时间提供了指导,以避免分散。 截至2012年6月,此修正案批准,并将被视为在韩国首尔的即将到来的ISO TC108会议期间作为最终修订。