Mechanical behavior, wear surface morphology, and clinical performance of UHMWPE acetabular components after 10 years of implantation

机译:植入10年后UHMWPE髋臼组件的力学行为,磨损表面形态和临床表现

获取原文

摘要

The relationships between the mechanical behavior of ultra-high molecular weight polyethylene (UHMWPE) and the long-term clinical performance of acetabular component bearings remain poorly understood. During previous hip simulator validation studies, researchers have correlated the mechanical behavior and surface morphology of polymeric biomaterials with the wear performance in a hip simulator. For the present study, we sought correlations between clinical performance and quantitative metrics of the mechanical behavior and surface morphology of retrieved cups. Nine all-UHMWPE acetabular cups of the same cemented design, 32 mm head size, and manufacturer were retrieved after an average implantation time of 14.1 years (range: 11.5-16.4 years). The implants were manufactured from RCH-1000, sterilized by gamma radiation in air and implanted between 1980 and 1983. Mechanical behavior of the retrieved components was determined using the previously validated small punch test, which subjects miniature specimens to multi-axial loading conditions. Surface morphology of the retrievals was assessed using white light interferometry. No significant relationship was observed between the surface roughness measurements and patient related variables. However, statistically significant relationships were observed between the mechanical behavior determined by the small punch test and the implantation time and patient weight associated with the retrieved hip cups. These findings support the hypothesis that the mechanical behavior of UHMWPE is related, at least in part, to the clinical performance of acetabular component bearings for total hip replacement.
机译:超高分子量聚乙烯(UHMWPE)的机械性能与髋臼组件轴承的长期临床性能之间的关系仍然知之甚少。在先前的髋关节模拟器验证研究中,研究人员已将聚合物生物材料的机械行为和表面形态与髋关节模拟器中的磨损性能相关联。对于本研究,我们寻求临床表现与回收杯的机械行为和表面形态的定量指标之间的相关性。平均植入时间为14.1年(范围:11.5-16.4年)后,取回了9个相同骨水泥设计的全UHMWPE髋臼杯,其头部尺寸为32 mm,且制造商相同。植入物是由RCH-1000制造的,并通过空气中的伽马射线辐射进行了灭菌,并于1980年至1983年之间植入。使用先前验证的小冲孔试验确定了回收部件的机械性能,该试验对微型样品进行了多轴加载条件的测试。使用白光干涉仪评估取回物的表面形态。在表面粗糙度测量值和患者相关变量之间未观察到显着关系。但是,在小冲床试验确定的机械行为与植入时间和与患者髋臼杯相关的患者体重之间,观察到统计学上显着的关系。这些发现支持以下假设:UHMWPE的机械行为至少部分与全髋关节置换髋臼组件轴承的临床性能有关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号