Abstract: The knowledge of where one has measured the response of a structure is sometimes as important as the response measured. This paper describes a method for the calibration of the horizontal and vertical scanners in a laser Doppler vibrometer. The residual of the scanner position with respect to a first-order and second-order regression model will be discussed. Comparisons will be made to vendor measured scanner characteristics. Detailed analysis of the scanner calibration data will result in a very accurate knowledge of where the measured data from a structure is located and directed. Moreover, this knowledge will enhance the accuracy of 3D laser vibrometry. !12
机译:摘要:有时人们对在何处测量结构的响应的了解与所测量的响应一样重要。本文介绍了一种用于校准激光多普勒振动计中水平和垂直扫描仪的方法。将讨论扫描仪位置相对于一阶和二阶回归模型的残差。将与供应商测量的扫描仪特性进行比较。对扫描仪校准数据的详细分析将非常准确地了解来自结构的测量数据的位置和方向。此外,该知识将提高3D激光振动计的准确性。 !12