In the last few years, the use of sentinel node technique has been increased; its success is connected with the proper performance of intraoperative gamma probes. This paper is focused on how to make performance and quality control tests to Europrobe S553U intraoperative gamma probe, from NEMA NU 3-2004 "Performance Measurement and Quality Control Guidelines for Non-Imaging Intraoperative Gamma Probes" normative, using devices designed for such purpose. Results of 11 performance tests are shown, which are between normal ranges previously reported in literature. Sensitivity values obtained in air are below of those obtained in a scatter medium, due to the collimation effect of the canon source. For all measured parameters, values with lower standard deviation belong to a distance of 30 mm between source and probe. Spatial resolution was (28,6±0,7)mm for a CdZnTe probe and (33,7±0,8)mm for a Csl one; best angular resolution in scatter medium was obtained for the first one of them (20,3±0,8degrees). A good energy resolution was obtained for CdZnTe detector (17kev of FWHM, for ~(99m)Tc). For periodic controls, an average activity of the punctual source around 0,74 MBq is suggested, in order to warranty a proper count rate, avoiding detectors saturation.
机译:在最近几年中,前哨节点技术的使用有所增加。它的成功与术中伽马探针的正确性能有关。本文的重点是如何使用为此目的设计的设备,根据NEMA NU 3-2004“非成像术中伽马射线探针的性能测量和质量控制指南”对Europrobe S553U术中伽马射线探针进行性能和质量控制测试。显示了11个性能测试的结果,该结果介于先前文献中报道的正常范围之间。由于佳能源的准直效应,在空气中获得的灵敏度值低于在散射介质中获得的灵敏度值。对于所有测得的参数,标准偏差较小的值在信号源和探头之间的距离为30 mm。 CdZnTe探针的空间分辨率为(28,6±0,7)mm,Csl探针的空间分辨率为(33,7±0.8)mm;对于它们中的第一个(20,3±0,8度),在散射介质中获得了最佳的角分辨率。对于CdZnTe检测器(FWHM的17kev,〜(99m)Tc)获得了良好的能量分辨率。对于定期控制,建议点源的平均活动约为0.74 MBq,以保证适当的计数率,避免检测器饱和。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号