Survey and Reduction of Patient Doses in Computed Tomography

机译:计算机断层扫描中患者剂量的调查和减少

获取原文

摘要

A survey of patient exposure in Computed Tomography was taken up by a team of Physicists from Dubai hospital in 2008 after the UAE joined the IAEA regional project on patient exposure in radiology, RAS/9/047. Nine major Hospitals from both public and private sectors in Dubai area were selected to participate to the survey. Phase 1 of the program consisted on checking the image quality of the CT machines in each Hospital and measuring the values of the CTDIW by means of the head (16 cm of diameter) and body (32 cm of diameter) cylindrical PMMA phantoms. After the conclusion of the first phase, Phase 2 was launched at Dubai Hospital, a major Governmental Hospital in Dubai, as a pilot study consisting of collecting dose-Length-Product (DLP) values generated by the CT system after CT examinations are completed. The examinations included in the study were brain, chest, and the combination Abdomen & Pelvis CT procedures, which are the most common in the Hospital. Furthermore, the recorded DLP values were compared to and found much higher than the Dose Reference Levels (DRL) established in the UK |Ref. 1|. The situation called for a remedial effort to reduce rapidly patient doses in Computed Tomography. The corrective actions undertaken jointly by Medical Physicists and radiology staff (the users) reached its objectives as the technical measures taken allowed a drastic reduction of the DLP values, bringing down patient exposure to acceptable levels, closer to the standards prevailing in international practice. The experience gained at Dubai Hospital is being implanted in the rest of the participating hospitals
机译:阿联酋加入原子能机构RAS / 9/047放射治疗区域项目后,阿联酋从2008年开始由迪拜医院的一组医生对计算机断层扫描中的患者暴露进行了调查。来自迪拜地区的9家来自公立和私立医院的大型医院被选中参加调查。该计划的第一阶段包括检查每个医院CT机的图像质量,并通过头部(直径16厘米)和身体(直径32厘米)圆柱形PMMA体模测量CTDIW的值。第一阶段结束后,第二阶段在迪拜医院(迪拜主要的政府医院)启动,作为一项试点研究,包括收集CT检查完成后由CT系统生成的剂量-长度-乘积(DLP)值。该研究中包括脑部,胸部以及腹部与骨盆CT联合检查程序,这是医院中最常见的检查。此外,将记录的DLP值与英国参考文献中建立的剂量参考水平(DRL)进行比较,并发现其高得多。 1 |。这种情况要求采取补救措施以减少计算机断层扫描中的快速患者剂量。医学物理学家和放射科工作人员(用户)共同采取的纠正措施达到了其目标,因为所采取的技术措施可以大大降低DLP值,使患者的暴露水平降低到可接受的水平,接近国际通用的标准。在迪拜医院获得的经验将被植入到其他参与医院中

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号