首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Tube Housing Leakiness Radiation and Patient Related Radiation Protective Measures - Gonads Radiation Exposure from Panoramic Radiography
【24h】

Tube Housing Leakiness Radiation and Patient Related Radiation Protective Measures - Gonads Radiation Exposure from Panoramic Radiography

机译:管壳泄漏辐射及与患者有关的辐射防护措施-全景射线照相的性腺辐射

获取原文

摘要

The purpose of this study was to compare patient- and apparatus related radiation protective measures considering gonad radiation exposure at panoramic radiography. The local dose (yGv) in the region of the gonads of a male Alderson-Rando-Phantom was measured by the dosimeter Tol-F (Berthold Technology, Germany) during panoramic radiography (PR) by the digital Orthophos XG plus*(Sirona / Bensheim, Germany). Different radiation protective measures was investigated during constant exposure parameters. In the second part of the survey the same ionisation chamber recorded the scattered radiation at the gonads level of the none protected phantom at various case shielding. Without protection apron the dose near to gonads amounted to 0,36 uGy in the mean (initial value), with ventrally applied X-ray apron 0,05 uGy (86% dosage reduction) and with ventral semi apron 0,025 uGy (93 % dosage reduction). At the investigation of tube housing leakiness radiation on the gonads level of the phantom 038 uGy was measured, at closed collimator (5mm lead) 0,36 uGy, at shielding the bottom of the apparatus 0,01 uGy. During a PR with opened Collimator (normal colli-mation) and additional basal shielding an equivalent dose of 0,03 uGy could be ascertained. The reinforced shielding at the bottom of the apparatus reduced the gonads radiation exposure to 92%. Concerning the gonadal protection the optimized patient related and the reinforced apparatus related radiation protection did not differ significantly. At Orthophos XG plus* the tube housing leakiness radiation is an significant factor of the nevertheless relative small radiation exposure of the gonads. The obvious reduction of the gonads exposure by ventral applying of apron or semi apron justifies this simple radiation protective precaution. An additional basal shielding on the apparatus fits the same way. At panoramic radiography the scattered radiation from the primary radiation beam is not significant determining the gonad dose.
机译:这项研究的目的是比较考虑全景射线照相中性腺辐射暴露的患者和设备相关的辐射防护措施。雄性Alderson-Rando-Phantom的性腺区域的局部剂量(yGv)在数字射线Orthophos XG plus *(Sirona /德国本斯海姆)。在恒定的暴露参数下研究了不同的辐射防护措施。在调查的第二部分中,同一电离室在各种情况下屏蔽时记录了未受保护的体模在性腺水平上的散射辐射。如果没有保护性围裙,则接近性腺的平均剂量(初始值)为0.36 uGy,腹侧使用X射线围裙的剂量为0.05 uGy(剂量减少86%),腹侧半围裙的剂量为0.025 uGy(93%剂量)减少)。在研究管壳泄漏时,在密闭式准直仪(5mm引线)上0.36 uGy处,在屏蔽装置底部0.01 uGy处,测量了幻影038 uGy的性腺水平上的辐射。在打开准直仪(正常准直)和附加基础屏蔽的PR期间,可以确定等效剂量为0.03 uGy。设备底部的增强屏蔽将性腺辐射暴露降低至92%。关于性腺保护,优化的患者相关和增强设备相关的辐射防护没有显着差异。在Orthophos XG plus *上,管壳泄漏辐射是性腺辐射暴露相对较小的重要因素。通过腹面或半围裙腹面明显减少性腺暴露,证明了这种简单的辐射防护措施是合理的。设备上的附加基础屏蔽层以相同的方式安装。在全景射线照相中,来自初级辐射束的散射辐射对决定性腺剂量并不重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号