Calorimetric Determination of Absorbed Dose to Water for an ~192Ir HDR Brachytherapy Source in Near-Field Geometry

机译:近场几何中约192Ir HDR近距离放射源的水吸收剂量的量热法测定

获取原文

摘要

~192Ir brachytherapy sources have been calibrated in terms of the reference air kerma rate at the Physikalisch-Technische Bundesanstalt (PTB) for more than 15 years. To reduce the uncertainty of the conversion factor from the reference air kerma rate to the absorbed dose to water at a distance of 1 cm from the source, the PTB has started a new project for the direct determination of the absorbed dose to water at a short distance from a clinical l92Ir HDR (high dose rate) source by use of a water calorimeter. The experimental set-up, heat transfer calculations and calorimetric measurements in three different depths in water are presented. For the determination of the dose rate constant A, additional measurements of the reference air kerma rate were performed and the radial dose function was determined by means of a plastic scintillation detector system as well as by Monte Carlo calculations.
机译:在Physikalisch-Technische Bundesanstalt(PTB)上已根据参考空气比释动能率校准了约192Ir近距离放射治疗源,已有15年以上的历史了。为了减少从参考空气比释动能率到距源的距离为1 cm的水的吸收剂量的转换因子的不确定性,PTB启动了一个新项目,可在短时间内直接确定对水的吸收剂量用水量热计距临床IrIrIr HDR(高剂量率)源的距离。介绍了在三个不同深度的水中的实验设置,传热计算和量热测量。为了确定剂量率常数A,进行了参考空气比释动能比的另外的测量,并且借助于塑料闪烁检测器系统以及通过蒙特卡洛计算来确定径向剂量函数。