首页> 外文会议> >Distortion of interaural time cues by directional noise reduction systems in modern digital hearing aids
【24h】

Distortion of interaural time cues by directional noise reduction systems in modern digital hearing aids

机译:现代数字助听器中的定向降噪系统使听觉时间提示失真

获取原文

摘要

In this paper we analyze the distortion of interaural time information by modern digital hearing aids. It is shown that the directionality in dual microphone hearing aids, created by a software directional microphone or an adaptive directional microphone, are very sensitive to intermicrophone mismatch, which in particular results in severe distortion of interaural time information. This interaural information is crucial for sound localization as well as for speech perception in noise.
机译:在本文中,我们分析了现代数字助听器对耳间时间信息的失真。结果表明,由软件定向麦克风或自适应定向麦克风创建的双麦克风助听器的方向性对麦克风间的不匹配非常敏感,特别是导致听觉时间信息的严重失真。这种听觉信息对于声音定位以及噪声中的语音感知都至关重要。