首页> 外文会议> >Advances in 2-D array Catheter transducers for real-time three-dimensional intracardiac Echocardiography
【24h】

Advances in 2-D array Catheter transducers for real-time three-dimensional intracardiac Echocardiography

机译:用于实时三维心内超声心动图的二维阵列导管换能器的进展

获取原文

摘要

In this paper, we describe advances in the design, fabrication and testing of two-dimensional catheter-mounted arrays for real-time three-dimensional (RT3-D) intracardiac echocardiography (ICE). These include a 7 Fr (2.3 mm O.D.), 112 element 2-D array utilizing a non-coaxial ribbon-based cabling technology. The 7 Fr transducer has a center frequency of 5 MHz, with a -6 dB fractional bandwidth of 30% and insertion loss of 80 dB (including 1 m of cable). The small size of the catheter enables RT3-D ICE from within the coronary sinus, a standpoint not possible with previous ICE catheters. We present in vivo sheep images of cardiac anatomy obtained from within the CS including the left and right atria, aorta, coronary arteries, and pulmonary veins. We also present simultaneous long and short axis views of the left ventricle, mitral valve, and left atrium with the 7 Fr catheter positioned in the right ventricle. We have also developed a 14 Fr (4.5 mm O.D.) forward-viewing 2-D array intracardiac probe containing an integrated working lumen for interventional device delivery. We present in vivo images of cardiac anatomy and simultaneous device delivery including the manipulation of a 0.35 mm diameter guidewire into the CS. We have also recently investigated the feasibility of using the forward-viewing probes to visualize the deployment of an abdominal aortic graft.
机译:在本文中,我们描述了用于实时三维(RT3-D)心内超声心动图(ICE)的二维导管安装阵列的设计,制造和测试方面的进展。其中包括利用非同轴带状电缆技术的7 Fr(外径2.3 mm),112元素2-D阵列。 7 Fr传感器的中心频率为5 MHz,-6 dB的分数带宽为30%,插入损耗为80 dB(包括1 m的电缆)。导管的小尺寸使得能够在冠状窦内进行RT3-D ICE,这是以前的ICE导管无法实现的观点。我们介绍了从CS内获得的心脏解剖结构的体内绵羊图像,包括左,右心房,主动脉,冠状动脉和肺静脉。我们还同时显示了左心室,二尖瓣和左心房的长轴和短轴视图,其中7 Fr导管位于右心室。我们还开发了一种14 Fr(O.D. 4.5 mm O.D.)前视二维阵列心内膜探头,该探头内含集成的工作腔,用于介入装置的输送。我们介绍了心脏解剖结构和同时进行的器械递送的体内图像,包括将直径为0.35毫米的导丝操纵入CS的操作。我们最近还研究了使用前视探头可视化腹主动脉移植物的部署的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号