The effect of grid misalignment on the uniformity of response of gas proportional scintillation counters

机译:栅格未对准对气体比例闪烁计数器响应均匀性的影响

获取原文

摘要

In the present work we revisit the simplifying assumptions of an earlier theoretical work on the effect of grid misalignnent on the performance of planar GPSCs. Experimental results that quantitatively establish this effect are presented. The magnitude of the effect is not as large as predicted due in part to the reduced dependence of the electroluminescence yield on the reduced electric field near threshold and due to electron cloud size. For the reduced fields normally used in the scintillation region of an optimized GPSC, misalignments as large as 3% have less than a 1% effect on pulse-amplitude variations. This effect is less than the statistical fluctuations in the number of primary electrons nominally produced in the absorption of X-rays. For reduced fields closer to the scintillation threshold, 15% variations in pulse amplitudes are possible, although it is unlikely that a GPSC would be operated under those conditions.
机译:在当前的工作中,我们将重新讨论早期的理论工作的简化假设,这些工作是关于网格未对准对平面GPSC性能的影响。提出了定量建立这种效果的实验结果。该影响的幅度没有预期的大,部分原因是由于电致发光产量对阈值附近的减小的电场的依赖性降低以及电子云的大小。对于通常用于优化GPSC闪烁区的缩小场,高达3%的失准对脉冲幅度变化的影响小于1%。这种影响小于在吸收X射线时名义上产生的一次电子数量的统计波动。对于更接近闪烁阈值的减小的场,脉冲幅度可能会发生15%的变化,尽管GPSC不太可能在这些条件下运行。