Contactless and Cuffless Monitoring of Blood Pressure on a Chair Using E-Textile Materials

机译:使用电子纺织材料的非接触式和无袖带式血压监测

获取原文

摘要

Blood pressure (BP) can be measured cufflessly and noninvasively by using pulse transit time, which is obtained from simultaneously recorded electrocardiogram (ECG) and photoplethysmogram (PPG). To apply this BP monitoring technique in daily life, an office chair was modified into a contactless and cuffless BP-monitoring device using e-textile materials. In this study, twelve normotensive subjects, including 6 males and 6 females, were recruited for testing this BP-monitoring chair. During the experiment, exercise was carried out to obtain a wider range of BP. In total, 62 pairs of systolic BP (SBP) and diastolic BP (DBP) estimations were made. Readings from a clinically approved cuff-based automatic BP monitor was used as reference. The results showed that the estimated SBP and DBP differed from their references by -6.3plusmn10.5 mmHg and -0.7plusmn4.9 mmHg respectively. It was found that the SBP estimation errors mainly came from the data obtained immediately after exercise. Further improvement should be made to minimize those errors. However, this concept is potentially useful in self BP monitoring for home healthcare.
机译:血压(BP)可以通过脉搏传输时间无袖且无创地测量,该时间可以从同时记录的心电图(ECG)和光体积描记图(PPG)获得。为了将这种BP监测技术应用到日常生活中,将办公椅改装为使用电子纺织材料的非接触式和无袖带BP监测设备。在这项研究中,招募了12名血压正常的受试者,包括6名男性和6名女性,以测试该BP监护椅。在实验过程中,进行了锻炼以获得更大范围的血压。总共进行了62对收缩压(SBP)和舒张压(DBP)估计。来自临床认可的基于袖带的自动BP监护仪的读数用作参考。结果表明,估计的SBP和DBP与参考值相差分别为-6.3plusmn10.5 mmHg和-0.7plusmn4.9 mmHg。发现SBP估计误差主要来自运动后立即获得的数据。应该进行进一步的改进以最大程度地减少这些错误。但是,此概念可能对家庭医疗保健的自我BP监视有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号