首页> 外文会议> >Quantitative CT characterization of body fluids with spectral 驴Z projection method
【24h】

Quantitative CT characterization of body fluids with spectral 驴Z projection method

机译:光谱驴projection投影法对体液进行CT定量表征。

获取原文
获取外文期刊封面目录资料

摘要

We have performed a clinical study on the characterization of body fluids with the spectral rhoZ projection algorithm. It converts dual-energy CT scans into density rho and atomic number Z information. We provide data on a total of 56 samples of 6 different classes. Blood, gall, pus and urine, as well as mixtures of blood + pus and blood + ethylendiamintetra acetate (EDTA, a blood anti-coagulant) have been investigated. For standard measurements with 80 kV and 140 kV tube voltages we encounter larger overlaps of the HU attenuation values. Gall fluid and pus are the only distinguishable substances. For the rhoZ projected image data, we find blood, pus and the blood + pus mixtures to be clearly distinguishable in the two-dimensional rhoZ plane. Gall fluid shows a small overlap to pus. The density and the atomic number values have standard deviations in the order of Deltarho = 10 mg/cm3 and DeltaZ = 0.1. The theoretical density and atomic number values for blood and water are reproduced with very small deviations in the range of Deltarho = -20 mg/cm3 and DeltaZ = 0.02.
机译:我们已经使用光谱rhoZ投影算法进行了体液表征的临床研究。它将双能CT扫描转换为密度rho和原子序数Z信息。我们提供了6种不同类别的总共56个样本的数据。已研究了血液,胆汁,脓液和尿液以及血液+脓液和血液+醋酸乙二胺四乙酸(EDTA,一种抗凝血剂)的混合物。对于使用80 kV和140 kV管电压的标准测量,我们会遇到HU衰减值的较大重叠。胆汁和脓液是唯一可区分的物质。对于rhoZ投影图像数据,我们发现血液,脓液和血液+脓液混合物在二维rhoZ平面中清晰可辨。胆汁显示出少量的脓液重叠。密度和原子序数值的标准偏差为Deltarho = 10 mg / cm 3 和DeltaZ = 0.1。复制的血液和水的理论密度和原子序数值在Deltarho = -20 mg / cm 3 和DeltaZ = 0.02的范围内有很小的偏差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号