Three detection algorithms used in implantable cardioverter defibrillators (ICDs) are described and compared. Synthetic and prerecorded human tachyarrhythmias were applied to algorithm simulations. As rhythm variability increased, the consistency of the result near detection zone boundaries fell and had different trends for each algorithm. For 456 prerecorded rhythms and fixed zones, VT-VF sensitivities were 96-97% and specificities were 83-85%.
机译:描述并比较了植入式心脏复律除颤器(ICD)中使用的三种检测算法。合成的和预先记录的人类快速性心律失常被用于算法模拟。随着节奏变化性的增加,检测区域边界附近的结果一致性下降,并且每种算法的趋势有所不同。对于456个预先录制的节律和固定区域,VT-VF敏感性为96-97%,特异性为83-85%。