A tritium detector for the Tokamak Fusion Test Reactor

机译:托卡马克聚变测试反应堆的detector探测器

获取原文

摘要

We have developed a device for on-line dynamic detection of tritium transferred during regeneration of neutral beam injector cryopanels at the Tokamak Fusion Test Reactor (TFTR) at Princeton Plasma Physics Laboratory (PPPL). The detector has been calibrated with various tritium-containing gas mixtures, dynamically tested in flowing gas streams and installed in TFTR's neutral beam forevacuum line. The device measures tritium partial pressure in the regeneration line by beta scintillation detection and total gas pressure by capacitance manometry. The tritium partial pressure detection range is 10/sup -5/ to 2 Torr-T/sub 2/, covered with three multiplier voltage gain settings. Careful operation, to avoid exposure to high tritium partial pressures, can lower the detection limit to 2/spl times/10/sup -6/ Torr of tritium. The relative error on tritium pressure measurement in a flowing mixture (1% T/sub 2/ in D/sub 2/) was less than 3%.
机译:我们开发了一种装置,用于在普林斯顿等离子体物理实验室(PPPL)的托卡马克聚变测试反应堆(TFTR)上在线动态检测在中性束注入器低温板再生过程中转移的tri。该检测器已经用各种含tri的气体混合物进行了校准,在流动的气流中进行了动态测试,并安装在TFTR的中性束真空管中。该设备通过β闪烁检测来测量再生管线中的partial分压,并通过电容测压法测量总气体压力。 t分压检测范围是10 / sup -5 /至2 Torr-T / sub 2 /,覆盖了三个倍增器电压增益设置。小心操作以避免暴露于high高压力,可将the的检测极限降低至2 / spl次/ 10 / sup -6 / Torr。流动混合物中(压力测量的相对误差(D / sub 2 /中为1%T / sub 2 /)小于3%。