Calorimeter with compensative absorber for measuring mu W level laser power

机译:带有补偿吸收器的量热仪,用于测量微瓦级激光功率

获取原文

摘要

A calorimeter with a compensative absorber developed for measuring mu W-level laser power is described. With this calorimeter, laser power measurements can be made irrespective of pressure fluctuations in the surroundings. The calorimeter was found to be very effective in reducing the effects of ambient pressure fluctuations. The pressure stability has been improved by more than tenfold, and the standard deviation of the measured value is 0.01 approximately 0.6% for 10-mW approximately 1- mu W power levels. The systematic error of the measurement, which is represented by the standard deviation (1 sigma ), was evaluated as +or-0.1% for 10 mu W.
机译:描述了一种带有补偿吸收器的量热仪,用于测量μW级激光功率。使用该热量计,可以进行激光功率测量,而与周围的压力波动无关。发现热量计对于减少环境压力波动的影响非常有效。压力稳定性提高了十倍以上,并且对于10 mW约1μW的功率水平,测量值的标准偏差为0.01约0.6%。测量的系统误差(以标准偏差(1 sigma)表示)在10μW时被评估为+或-0.1%。