Abstract: In this paper, we evaluate our ability to achieve consistent image presentation across a wide range of output devices, focusing on digital x-ray radiography for chest applications. In particular we focus on dry versus wet printers of hardcopy prints. In this evaluation, we review the expected theoretical variability using the DICOM grayscale standard display function (GSDF). The GSDF maps DICOM presentation values to luminance values that are perceived by a human. We present our methodology for calibrating devices as evaluated on sixteen printers. Seven devices were selected for a human observer study to determine if there are perceptible differences in the presentation of a given image, focusing on differences between wet and dry processes. It was found that wet printers were preferred, however, there may be other logistical and practical reasons whey dry printers may be used. !4
机译:摘要:在本文中,我们评估了在广泛的输出设备上实现一致图像呈现的能力,重点是针对胸部应用的数字X射线摄影。特别是,我们专注于硬拷贝打印的干式和湿式打印机。在此评估中,我们使用DICOM灰度标准显示功能(GSDF)审查了预期的理论变异性。 GSDF将DICOM表示值映射到人类可以感知的亮度值。我们介绍了在16台打印机上评估过的校准设备的方法。选择了七个用于人类观察者研究的设备,以确定给定图像的呈现方式是否存在可感知的差异,重点是干湿法之间的差异。已经发现湿打印机是优选的,但是,在物流和实践上还有其他原因可以使用乳清干打印机。 !4