Design guidelines for medical ultrasonic arrays

机译:医用超声阵列设计指南

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The basics of ultrasonic transducer array design in the frequency range useful for medical imaging are discussed. Performance parameters of importance in transducer design are considered, including sensitivity, coupling constant, band width, frequency downshift, pulse duration, beam focusing properties, and electrical matching. 2D and 3D effects must also be taken into account. The advantages of computer modeling in 3D with finite element analysis code are highlighted. The principles of multi-element array transducers useful for 2D real-time scanning are reviewed. 2D arrays provide the opportunity of focusing the elevation beam dimension in the short axis, or make possible full 3D scanning volumes. !54
机译:摘要:讨论了可用于医学成像的频率范围内超声换能器阵列设计的基础。考虑了换能器设计中重要的性能参数,包括灵敏度,耦合常数,带宽,降频,脉冲持续时间,束聚焦特性和电匹配。还必须考虑2D和3D效果。强调了使用有限元分析代码进行3D计算机建模的优点。回顾了对二维实时扫描有用的多元素阵列换能器的原理。 2D阵列提供了在短轴上聚焦高程光束尺寸的机会,或者使完整的3D扫描体积成为可能。 !54