首页> 外文会议>Medical applications of radiation detectors III >Dark-count-less X-ray photon counting using an LSO-MPPC detector and its application to computed tomography
【24h】

Dark-count-less X-ray photon counting using an LSO-MPPC detector and its application to computed tomography

机译:LSO-MPPC检测器的无暗数X射线光子计数及其在计算机断层摄影中的应用

获取原文
获取原文并翻译 | 示例

摘要

X-ray photons are detected using an Lu_2(SiO_4)O [LSO] single-crystal scintillator with a decay time of 40 ns and a multipixel photon counter (MPPC). The photocurrent from the MPPC is amplified by a high-speed current-voltage amplifier with an 80 MHz-gain-band operational amplifier, and the 200-ns-width event pulses are sent to a multichannel analyzer to measure X-ray spectra. The MPPC is driven in the pre-Geiger mode at a bias voltage of 70.7 V and a temperature of 23℃. Photon-counting computed tomography (PC-CT) is accomplished by repeated linear scans and rotations of an object, and projection curves of the object are obtained by linear scanning. The exposure time for obtaining a tomogram was 10 min with scan steps of 0.5 mm and rotation steps of 1.0°. At a tube voltage of 100 kV, the maximum count rate was 350 kcps/pixel. We carried out PC-CT using gadolinium media and confirmed the energy-dispersive effect with changes in the lower level voltage of event pulses using a comparator.
机译:使用Lu_2(SiO_4)O [LSO]单晶闪烁体(衰减时间为40 ns)和多像素光子计数器(MPPC)检测X射线光子。来自MPPC的光电流由具有80 MHz增益频带运算放大器的高速电流-电压放大器放大,并将200 ns宽度的事件脉冲发送到多通道分析仪以测量X射线光谱。 MPPC在预盖革模式下以70.7 V的偏置电压和23℃的温度驱动。光子计数计算机断层扫描(PC-CT)通过重复进行线性扫描和对象旋转来实现,并且通过线性扫描获得对象的投影曲线。获得断层图像的曝光时间为10分钟,扫描步长为0.5 mm,旋转步长为1.0°。在100 kV的管电压下,最大计数率为350 kcps /像素。我们使用g介质进行了PC-CT,并使用比较器确认了事件脉冲的低电平电压变化时的能量分散效应。

著录项

  • 来源
    《Medical applications of radiation detectors III》|2013年|885302.1-885302.5|共5页
  • 会议地点 San Diego CA(US)
  • 作者单位

    Department of Physics, Iwate Medical University, 2-1-1 Nishitokuta, Yahaba, Iwate 028-3694,Japan;

    Department of Physics, Iwate Medical University, 2-1-1 Nishitokuta, Yahaba, Iwate 028-3694,Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    The 3rd Department of Surgery, Toho University School of Medicine, 2-17-6 Ohashi, Meguro,Tokyo 153-8515, Japan;

    Department of Microbiology, School of Medicine, Iwate Medical University, 2-1-1 Nishitokuta,Yahaba, Iwate 028-3694, Japan;

    Department of Neurosurgery, School of Medicine, Iwate Medical University, 19-1 Uchimaru,Morioka, Iwate 020-0023, Japan;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    LSO-MPPC detector; dark-count-less; pre-Geiger mode; X-ray spectra; X-ray CT; energy-dispersive effect; gadolinium imaging;

    机译:LSO-MPPC检测器;不计其数预盖革模式X射线光谱; X射线CT能量分散效应; imaging成像;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号