On Measurement Uncertainty of a Vortex Flowmeter

机译:涡街流量计的测量不确定度

获取原文
获取原文并翻译 | 示例

摘要

Measurement uncertainty of a T-shaped flowmeter was examined with a gas flow calibration facility, for Reynolds numbers, Re_d, in a range of 2.56x10~4-1.56x10~5. The total uncertainty interval of the Strouhal values deduced from a Fourier spectral analysis was ±0.745%, and the linearity of the data was ±1.78%. Meanwhile, a scheme based on pulses counting in reference to the analog signals measured gave the total uncertainty interval of the meter factor as ±0.568%, with the linearity ±1.89%.
机译:使用气体流量校准设备检查了T型流量计的测量不确定度,雷诺数Re_d的范围为2.56x10〜4-1.56x10〜5。傅里叶光谱分析得出的Strouhal值的总不确定区间为±0.745%,数据的线性度为±1.78%。同时,基于参考测量的模拟信号的脉冲计数的方案给出了仪表系数的总不确定度间隔为±0.568%,线性度为±1.89%。