ANALYSIS OF ERRORS OF COAL QUALITY MONITORS

机译:煤质监测仪误差分析

获取原文

摘要

On-line nuclear meters have been in use in the coal industry for many years. They have been utilised for coal quality monitoring, in the control systems for coal blending, or for separating coals in the heavy media separation process. Their operation is based on the scattering or the absorption of incident gamma radiation, and the derived density or ash value is the result of a time-averaged measurement. In this paper, errors of ash monitors have been analysed. These errors appear due to several reasons: measurements are performed on samples which inaccurately represent a whole mass of tested coal, measuring methods are sensitive to variations in chemical composition of ash, size distribution of coal and moisture content. The signal from the radiation detector has to be averaged over a long period of time which results in delay of measuring results.
机译:在线核电表已经在煤炭行业中使用了很多年。它们已用于煤炭质量监控,配煤控制系统或重介质分离过程中的煤炭分离。它们的操作基于入射伽马射线的散射或吸收,得出的密度或灰分值是时间平均测量的结果。本文对灰分监测器的误差进行了分析。这些错误的出现是由于以下几个原因:对不准确代表被测煤总量的样品进行测量,测量方法对灰分化学成分,煤的粒度分布和水分含量的变化敏感。来自辐射探测器的信号必须经过长时间的平均,这会导致测量结果的延迟。