首页> 外文会议>IEEE International Symposium on Electromagnetic Compatibility;IEEE Asia-Pacific Symposium on Electromagnetic Compatibility >In-vitro measurement system for induced voltages in medical implants in magnetic resonance imaging
【24h】

In-vitro measurement system for induced voltages in medical implants in magnetic resonance imaging

机译:磁共振成像中医用植入物中感应电压的体外测量系统

获取原文
获取外文期刊封面目录资料

摘要

Certification of active medical implants for use in magnetic resonance imaging machines requires the testing of immunity with respect to radiofrequency voltages induced into the active electronics of the implants. The testing of immunity is typically performed on the bench using an injection circuitry. However, the injection test voltage must be determined under real exposure conditions as they would occur in real human inside MRI. We have developed an optical differential voltage probe to be used in electromagnetically hostile environments like magnetic resonance imaging that is able to deliver the full complex-valued voltage signal with a dynamic range of 120 dB and a frequency range from 100 kHz to 1 GHz and with a total measurement uncertainty of 1.14 dB.
机译:核磁共振成像仪中使用的有源医疗植入物的认证要求测试针对植入到植入物的有源电子器件中的射频电压的抗扰度。免疫力测试通常是使用注入电路在工作台上进行的。但是,必须在真实的​​暴露条件下确定注入测试电压,因为它们会在MRI内部的真实人体中发生。我们已经开发出一种光差分电压探头,可用于电磁干扰环境(如磁共振成像)中,该探头能够提供动态范围为120 dB,频率范围为100 kHz至1 GHz的全复数值电压信号。总测量不确定度为1.14 dB。