Use of quality assurance procedures for FTIR field measurement data

机译:对FTIR现场测量数据使用质量保证程序

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The OP-FTIR measurement technique and other remote sensing systems (for example UV-DOAS and TDL) become more and more established methods to monitor the ambient air quality. Some quality assurance procedures have to be done on the field measurement data to get more reliable results. Both the German and US-EPA standardization documents for the OP-FTIR technique prescribe such procedures to check the data. To assess the applicability of these procedures, OP-FTIR field measurement data of a three month continuous measurement campaign were evaluated concerning the signal strength, the influence of stray light, water vapor concentration, baseline noise, wave number shifts and the detection limits. In addition the long time signal stability of the OP-FTIR systems and the UV-DOAS systems are intercompared. The results of the several quality assurance procedures will be presented in this paper. !19
机译:摘要:OP-FTIR测量技术和其他遥感系统(例如UV-DOAS和TDL)成为监控环境空气质量的越来越成熟的方法。为了获得更可靠的结果,必须对现场测量数据执行一些质量保证程序。关于OP-FTIR技术的德国和US-EPA标准化文件都规定了检查数据的程序。为了评估这些程序的适用性,对三个月连续测量活动的OP-FTIR现场测量数据进行了评估,涉及信号强度,杂散光的影响,水蒸气浓度,基线噪声,波数偏移和检测极限。另外,还比较了OP-FTIR系统和UV-DOAS系统的长时间信号稳定性。本文将介绍几种质量保证程序的结果。 !19