Introduction of argon beam coagulation functionality to robotic procedures using the ABC D-Flex Probe: equivalency to an existing laparoscopic instrument

机译:使用ABC D-Flex探针将氩气凝结功能引入机器人程序:与现有的腹腔镜仪器相当

获取原文
获取原文并翻译 | 示例

摘要

Introduction: The ABC D-Flex Probe utilizes argon beam coagulation (ABC) technology to achieve hemostasis during minimally invasive surgery. A handle on the probe allows for integration with robotic surgical systems and introduces ABC to the robotic toolbox. To better understand the utility of D-Flex, this study compares the performance of the D-Flex probe to an existing ABC Laparoscopic Probe through ex-vivo tissue analysis. METHODS: Comparisons were performed to determine the effect of four parameters: ABC device, tissue type, activation duration, and distance from tissue. 10 ABC D-Flex probes were used to create 30 burn samples for each comparison. Ex-vivo bovine liver and porcine muscle were used as tissue models. The area and depth of each burn was measured using a light microscope. The resulting dimensional data was used to correlate tissue effect with each variable. RESULTS: D-Flex created burns which were smaller in surface area than the Laparoscopic probe at all power levels. Additionally, D-Flex achieved thermal penetration levels equivalent to the Laparoscopic probe. CONCLUSION: D-Flex implements a small 7F geometry which creates a more focused beam. When used with robotic precision, quick localized superficial hemostasis can be achieved with minimal collateral damage. Additionally, D-Flex achieved equivalent thermal penetration levels at lower power and argon flow-rate settings than the Laparoscopic probe.
机译:简介:ABC D-Flex探针利用氩束凝结(ABC)技术在微创手术中实现止血。探头上的手柄可与机器人手术系统集成,并将ABC引入机器人工具箱。为了更好地了解D-Flex的效用,本研究通过离体组织分析将D-Flex探头的性能与现有的ABC腹腔镜探头进行了比较。方法:进行比较以确定四个参数的效果:ABC设备,组织类型,激活持续时间和与组织的距离。每个比较使用10个ABC D-Flex探针创建30个烧伤样品。离体牛肝和猪肌肉用作组织模型。使用光学显微镜测量每次烧伤的面积和深度。所得的尺寸数据用于将组织效果与每个变量相关联。结果:在所有功率水平下,D-Flex产生的烧伤表面积均小于腹腔镜探头。此外,D-Flex的热渗透水平相当于腹腔镜探头。结论:D-Flex实现了小的7F几何形状,从而产生了更聚焦的光束。当以机械手的精度使用时,可以实现快速的局部浅表止血,而对附带的损害最小。此外,与腹腔镜探头相比,D-Flex在更低的功率和氩气流量设置下达到了等效的热渗透水平。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号