Previous results revealed that the influences, caused by both individual differences and measuring environmental factors, would be reduced by using the floating-reference theory (FRT) for diffused reflectance spectra (DRS) -based blood glucose measurement by Near Infrared Spectroscopy (NIRS). The obtained signals could highlight the variation in light intensity which was brought only by the change of glucose concentration. The existing studies on FRT have mainly focused on the diffused reflectance spectra, but rarely involved the diffused transmittance spectra. In this talk, it is our aim to investigate the availability of FRT on the diffused transmittance spectra (DTS) on the basis of Monte Carlo (MC) simulation method. The MC simulations of DTS have been carried out with different glucose concentrations and skin tissue thicknesses. The simulation results show that the floating reference position point of DTS will disappear when the tissue thickness is greater than a certain value. Therefore, the FRT might be applied on thin tissue model for DTS by NIRS.
机译:先前的结果表明,通过使用基于近红外光谱(NIRS)的基于漫反射光谱(DRS)的血糖测量的浮动参考理论(FRT),可以减少由个体差异和测量环境因素引起的影响。所获得的信号可以突出仅由葡萄糖浓度的变化带来的光强度的变化。现有的关于FRT的研究主要集中在漫反射光谱上,但是很少涉及漫透射光谱。在本次演讲中,我们的目的是基于蒙特卡罗(MC)模拟方法研究在漫透射光谱(DTS)上FRT的可用性。 DTS的MC模拟是在不同的葡萄糖浓度和皮肤组织厚度下进行的。仿真结果表明,当组织厚度大于一定值时,DTS的浮动参考位置点将消失。因此,FIR可能被NIRS应用于DTS的薄组织模型。

著录项

  • 来源
  • 会议地点 San Francisco CA(US)
  • 作者单位

    Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, College of Precision Instruments Opto-electronics Engineering, Tianjin University, Tianjin 300072, China;

    Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, College of Precision Instruments Opto-electronics Engineering, Tianjin University, Tianjin 300072, China;

    Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, College of Precision Instruments Opto-electronics Engineering, Tianjin University, Tianjin 300072, China;

    Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, College of Precision Instruments Opto-electronics Engineering, Tianjin University, Tianjin 300072, China;

    Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, College of Precision Instruments Opto-electronics Engineering, Tianjin University, Tianjin 300072, China;

    State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    floating-reference theory; the availability; diffused transmittance; diffused reflectance; Monte Carlo simulation;

    机译:浮动参照理论可用性;漫透射率漫反射率蒙特卡洛模拟;
  • 入库时间 2022-08-26 14:31:16

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号