Spectrum Correction Considering Light Source Fluctuation for Non-invasive Blood Glucose Sensing

机译:考虑光源波动的频谱校正,用于非侵入性血糖传感

获取原文
获取原文并翻译 | 示例

摘要

The purpose of this study is to correct baseline shift in absorbance spectrums caused by light source fluctuation. To improve quantitative evaluation performance of blood glucose level, baseline shift is corrected by multiple scatter correction (MSC). Moreover, to increase the effect of the MSC, water vapor absorbance is subtracted, and relative glucose absorbance are calculated by dividing with hemoglobin absorbance at 1544 [cm-1]. In order to verify the effectiveness of the proposed spectrum correction method, light source fluctuation is simulated on the Fourier transform infrared spectroscopy (FT-IR), and we apply the proposed method to the spectrums measured by FT-IR. From the simulation results, the baseline shift was successfully reduced by proposed method.
机译:这项研究的目的是纠正由光源波动引起的吸收光谱的基线偏移。为了提高血糖水平的定量评估性能,通过多次散射校正(MSC)校正基线偏移。此外,为了增加MSC的作用,减去水蒸气吸收量,并通过除以1544 [cm-1]的血红蛋白吸收量来计算相对葡萄糖吸收量。为了验证所提出的光谱校正方法的有效性,在傅立叶变换红外光谱(FT-IR)上模拟了光源波动,并将该方法应用于通过FT-IR测量的光谱。从仿真结果来看,通过提出的方法成功地减少了基线偏移。