【24h】

Evaluation of hand-held gamma camera with 1-mm NaI(Tl) pixels

机译:具有1毫米NaI(Tl)像素的手持式伽马相机的评估

获取原文
获取原文并翻译 | 示例

摘要

A high resolution, hand-held gamma camera has been constructed for use in pre-surgical and intra-operative lymphoscintigraphy. In this paper, we evaluate a compact gamma camera system utilizing NaI(Tl) as a more likely candidate for scintillator due to its greater light yield and faster decay time than CsI(Na) used in an earlier prototype. Using NaI(Tl), the system mean energy resolution is 13% FWHM at 122keV, as compared to 28% in the system with CsI(Na). The highly compact detector head has a 2-cm by 2-cm field of view (FOV) and 1.25-mm intrinsic spatial resolution. Sensitivity of the NaI(Tl) camera was compared with the CsI(Na) camera. At 1cm distance from the tip of collimator, pinhole sensitivity was 58cps/μCi for the Na(Tl) camera and 37cps/μCi for the CsI(Na) camera.
机译:高分辨率的手持式伽马相机已被构造用于手术前和术中淋巴闪烁显像。在本文中,我们评估了一种紧凑的伽马相机系统,该系统利用NaI(Tl)作为闪烁体的可能性更高,因为与早期原型中使用的CsI(Na)相比,它具有更高的光输出量和更快的衰减时间。使用NaI(Tl),系统平均能量分辨率为122keV,FWHM为13%,而使用CsI(Na)的系统的平均能量分辨率为28%。高度紧凑的探测器头具有2厘米乘2厘米的视野(FOV)和1.25毫米的固有空间分辨率。将NaI(Tl)相机的灵敏度与CsI(Na)相机进行了比较。在距准直仪尖端1cm处,Na(Tl)相机的针孔灵敏度为58cps /μCi,而CsI(Na)相机的针孔灵敏度为37cps /μCi。