Hard tissue ablation with sub-mus CO_2 laser pulses with the use of an air-water spray

机译:使用空气-水喷雾用亚微米CO_2激光脉冲对硬组织进行消融

获取原文
获取原文并翻译 | 示例

摘要

Using short 300 ns pulses of a Q-switch CO_2 laser at the wavelength fo maximal hard tissue absorption (9.6 mum) we achieve charring-free ablation of bone tissue. An air-water spray prevents tissue parching and helps to avoid excessive rest-heat accumulation. The observed thermally altered layer at the cut surface in a cortical bone is of only 2-6 mum thickness. A long enough irradiation results in a progressively narrowing tot he bottom wedge-shaped cut profile and in a drop of the ablation rate as compared to its initial value. The main reason for this is work likely an enhancement of a heat dispersion and growing light absorption by ablation products. An enlargement of the cut width helps to avoid these negative phenomena. Using this technique we overcome restriction on the cut depth and reduce noticeably the cut time. With 66 W of average power from the Q-switch CO_2 laser we need 1/2 min to produce a 6-mm deep and 10-mm long in vitro incision in a hard cortical bone (young bull femur). That time increases to 2 min for 10-mm depth.
机译:使用Q开关CO_2激光的短300 ns脉冲,最大吸收硬组织(9.6毫米)的波长,我们实现了无烧伤的骨组织消融。空气喷水可防止组织干裂,并有助于避免过多的余热积聚。在皮骨切面处观察到的热变层厚度只有2-6微米。足够长的照射导致底部楔形切口轮廓逐渐变窄,并且与初始值相比,烧蚀率下降。造成这种现象的主要原因是可能通过消融产品增强了散热并增加了光吸收。切口宽度的扩大有助于避免这些不利现象。使用这种技术,我们克服了对切割深度的限制,并显着减少了切割时间。利用Q开关CO_2激光器的平均功率为66 W,我们需要1/2分钟的时间在硬皮层骨(年轻股骨)中产生6毫米深,10毫米长的体外切口。对于10毫米深度,该时间增加到2分钟。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号