Differential Phase Contrast Imaging (DPCI) has the potential to vastly increase soft tissue contrast. DPCI requires spatial and temporal coherence as generated by a synchrotron or a micro-focus X-ray source; however, recent research demonstrates DPCI can be implemented using a conventional X-ray source with three transmission gratings (Pfeiffer et al., Nature 2006). This paper describes the optimization of the essential system parameters (system size, delivered dose, spatial resolution) of this implementation from a theoretical perspective. The optimization of these parameters is an essential step in practical application of DPCI. We conclude that the minimum size of the system is approximately 700 mm, the minimum resolution is 100 um, and the dose is 1/1000 that of conventional absorption CT.
机译:差分相差成像(DPCI)有可能极大地增加软组织的对比度。 DPCI需要由同步加速器或微焦点X射线源生成的空间和时间相干性;但是,最近的研究表明,可以使用具有三个透射光栅的常规X射线源来实现DPCI(Pfeiffer等人,Nature 2006)。本文从理论的角度描述了此实施方案的基本系统参数(系统尺寸,输送剂量,空间分辨率)的优化。这些参数的优化是DPCI实际应用中必不可少的步骤。我们得出的结论是,系统的最小尺寸约为700 mm,最小分辨率为100 um,剂量为常规吸收CT的1/1000。