首页> 外文会议>Conference on Earth Observing Systems VII, Jul 7-10, 2002, Seattle, USA >Aqua MODIS Thermal Emissive Bands Calibration Algorithm and Preliminary Results
【24h】

Aqua MODIS Thermal Emissive Bands Calibration Algorithm and Preliminary Results

机译:Aqua MODIS热发射带校准算法和初步结果

获取原文
获取原文并翻译 | 示例

摘要

The MODerate Resolution Imaging Spectroradiometer (MODIS) Flight Model (FM1) was launched on-board the NASA Earth Observing System (EOS) Aqua spacecraft on May 4, 2002. Following initial instrument outgas, focal plane cooling, and functional testing, Aqua MODIS opened its nadir aperture door (NAD) on June 24, 2002 and acquired the first on-orbit image. MODIS has 36 spectral bands with wavelengths ranging from 0.412μ to 14.5μ. The 16 thermal emissive bands (B20-25 and 27-36) with wavelengths above 3.5μare calibrated on-orbit by an on-board calibrator blackbody (OBC BB) on a scan by scan basis. In this paper, we provide a brief review of MODIS thermal emissive bands (TEB) calibration algorithm and present early results from Aqua MODIS on-orbit calibration and characterization, including sensor's response stability, noise assessment, and the response versus scan angle (RVS) difference between the two sides of the scan mirror. In general, the on-orbit performance of Aqua MODIS TEB has been consistent with the instrument pre-launch calibration and characterization. Compared to its Protoflight Model (PFM), or the Terra MODIS, launched on December 18, 1999, a number of improvements have been achieved. During instrument on-orbit operation, the BB can be operated from 270-315K (during scheduled warm-up/cool-down cycle). When the BB temperature is above certain limits, three thermal emissive bands (33, 35, and 36) saturate when directly viewing the calibration source. We have developed an algorithm of using the Look-Up Table (LUT) gain coefficients for these three bands to keep the Earth view data calibrated under this situation. An illustration of on-orbit performance of this approach is provided.
机译:MODerate分辨率成像光谱仪(MODIS)飞行模型(FM1)于2002年5月4日在NASA地球观测系统(EOS)Aqua航天飞机上发射。在最初的仪器排气,焦平面冷却和功能测试之后,Aqua MODIS开了它的天底孔径门(NAD)于2002年6月24日获得了第一张在轨图像。 MODIS具有36个光谱带,波长范围从0.412μ至14.5μ。波长超过3.5μ的16个热发射带(B20-25和27-36)由机载校准器黑体(OBC BB)在轨道上逐次扫描校准。在本文中,我们简要介绍了MODIS热发射带(TEB)校准算法,并提供了Aqua Aqua的在轨校准和表征的早期结果,包括传感器的响应稳定性,噪声评估以及响应与扫描角度(RVS)扫描镜两侧之间的差异。通常,Aqua MODIS TEB的在轨性能与仪器发射前的校准和表征一致。与1999年12月18日推出的Protoflight模型(PFM)或Terra MODIS相比,已经实现了许多改进。在仪器在轨运行期间,BB可以在270-315K的范围内运行(在预定的预热/冷却周期内)。当BB温度高于特定限制时,直接查看校准源时,三个热发射带(33、35和36)会饱和。我们已经开发了一种算法,使用这三个频段的查找表(LUT)增益系数来在这种情况下保持对地球视图数据的校准。提供了这种方法在轨性能的说明。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号