An essential aspect for the neurological examination is the assessment of the deep tendon reflex. The deep tendon reflex is commonly conducted by a clinician through evoking the patellar tendon reflex. The significant parameters are the reflex response and the latency of the reflex loop. The characteristics of the reflex response and latency are representative of the severity and status attributes for upper motor neuron syndrome and peripheral neuropathy. The qualitative assessment from a clinician can be quantified through an ordinal scale, for example the NINDS Myotatic Reflex Scale. The accuracy of the ordinal scales for reflex response quantification is a subject of controversy. Another issue is the ordinal scales do not include temporal characterization of the reflex response. Latency can be derived with electrodiagnostic examination. However, electrodiagnostic examinations inherently involve the use of specialized and expensive resources. An alternative system that integrates tandem operated wireless 3D MEMS accelerometers for the quantification of reflex response and latency has been developed. The reflex response is quantified through positioning a wireless 3D MEMS accelerometer on the lower leg. A tandem and synchronized wireless 3D MEMS accelerometer is mounted to the swing arm that provides a quantified input for evoking the reflex. Latency is derived from the temporal disparity of the synchronized acceleration waveforms from the tandem wireless 3D MEMS accelerometers. Given the inherent variability of the deep tendon reflex, a mechanical artificial reflex device has been developed to evaluate the reflex quantification system incorporating wireless 3D MEMS accelerometers. The test and evaluation of the wireless 3D MEMS accelerometer reflex quantification system demonstrates considerable precision and accuracy imperative for reproducibility.
机译:神经系统检查的一个重要方面是深腱反射的评估。深肌腱反射通常由临床医生通过引起pa腱反射来进行。重要参数是反射反应和反射循环的潜伏期。反射反应和潜伏期的特征代表了上运动神经元综合征和周围神经病变的严重程度和状态属性。临床医生的定性评估可以通过序数量表(例如NINDS肌动反射量表)进行量化。反射反应量化的序数刻度的准确性是一个有争议的主题。另一个问题是序数标度不包括反射反应的时间特征。延迟可通过电诊断检查得出。但是,电诊断检查本质上涉及使用专门且昂贵的资源。已经开发了一种替代系统,该系统集成了串联操作的无线3D MEMS加速度计,用于量化反射响应和等待时间。通过将无线3D MEMS加速度计放在小腿上,可以量化反射响应。串联和同步的无线3D MEMS加速度计安装在摇臂上,为量化反射提供了量化的输入。延迟是从串联无线3D MEMS加速度计的同步加速度波形的时间差异得出的。考虑到深层肌腱反射的固有可变性,已经开发了一种机械人工反射设备来评估结合了无线3D MEMS加速度计的反射量化系统。无线3D MEMS加速度计反射定量系统的测试和评估证明了可重复性具有相当高的精度和准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号