摘要

Optical Coherence Tomography (OCT) is a relatively new type of imaging system for medical diagnosis. Because most current OCT systems use a sharply focused beam in tissues, they have a short depth of field (high image resolution is near the focus only). In this paper, limited diffraction beams of different orders are used to increase depth of field and to reduce sidelobes in OCT. Results show that the proposed OCT system has a lateral resolution of about 4.4 wavelengths (the central wavelength of the source is about 940 nm with a bandwidth of about 70 nm) and lower than -60 dB sidelobes over an entire depth of field of 4.5 mm with the diameter of the objective lens of 1 mm.
机译:光学相干断层扫描(OCT)是一种用于医学诊断的相对新型的成像系统。因为大多数当前的OCT系统在组织中使用聚焦强烈的光束,所以它们的景深很短(高图像分辨率仅在焦点附近)。在本文中,使用不同阶次的有限衍射光束来增加景深并减少OCT中的旁瓣。结果表明,提出的OCT系统的横向分辨率约为4.4个波长(光源的中心波长约为940 nm,带宽约为70 nm),并且在4.5 mm的整个景深中旁瓣低于-60 dB物镜的直径为1毫米。