Detection of early caries by laser-induced breakdown spectroscopy

机译:用激光诱导击穿光谱法检测早期龋齿

获取原文
获取原文并翻译 | 示例

摘要

To improve sensitivity of dental caries detection by laser-induced breakdown spectroscopy (LIBS) analysis, it is proposed to utilize emission peaks in the ultraviolet. We newly focused on zinc whose emission peaks exist in ultraviolet because zinc exists at high concentration in the outer layer of enamel. It was shown that by using ratios between heights of an emission peak of Zn and that of Ca, the detection sensitivity and stability are largely improved. It was also shown that early caries are differentiated from healthy part by properly setting a threshold in the detected ratios. The proposed caries detection system can be applied to dental laser systems such as ones based on Er:YAG-lasers. When ablating early caries part by laser light, the system notices the dentist that the ablation of caries part is finished. We also show the intensity of emission peaks of zinc decreased with ablation with Er:YAG laser light.
机译:为了提高通过激光诱导击穿光谱法(LIBS)分析龋齿检测的灵敏度,提出了利用紫外线中的发射峰。由于锌以高浓度存在于牙釉质的外层中,因此我们将重点放在发射峰存在于紫外线中的锌。结果表明,通过使用Zn和Ca的发射峰的高度之比,检测灵敏度和稳定性大大提高。还显示出通过适当设置检测比率的阈值,可以使早期龋病与健康部位区分开。所提出的龋齿检测系统可以应用于牙科激光系统,例如基于Er:YAG激光的系统。当用激光消融早期龋齿部分时,系统会通知牙医,消融龋齿部分已完成。我们还显示,随着Er:YAG激光烧蚀,锌的发射峰强度降低。