BIOSENSOR BASED ON CHEMILUMINESCENCE FOR SERUM URIC ACID DETERMINATION

机译:基于化学发光的生物传感器测定血清尿酸

获取原文
获取原文并翻译 | 示例

摘要

A biosensor based on luminol chemiluminescent reaction and immobilized uricase column for uric acid determination has been developed. The response time by the sensor for various concentration of serum uric acid was 47 seconds with 17 μl sample volume and at 40 samples per hour. The linear range of standard curve was from 1 mg/dl up to 20 mg/dl of serum uric acid. The within day imprecision was 3.22-4.36%. The day to day imprecision was 6.18-7.80%. The recovery rate by the method was 93-109%. Compared with standard co-lorimetric method employed enzymatic kit revealed that the linear regression and correlation coefficient were Y=0.4+0.938X and r=0.9909 respectively. The immobilized uricase column retained 94% of its original activity even after over 2000 runs for five and half monthes' continual usage.
机译:开发了一种基于鲁米诺化学发光反应和固定化尿酸酶柱的生物传感器,用于尿酸测定。传感器对各种浓度的血清尿酸的响应时间为47秒,样品体积为17μl,每小时40个样品。标准曲线的线性范围为1 mg / dl至20 mg / dl的血清尿酸。日内不精确度为3.22-4.36%。每天的不准确性为6.18-7.80%。该方法的回收率为93-109%。与标准的比色法相比,酶试剂盒的线性回归和相关系数分别为Y = 0.4 + 0.938X和r = 0.9909。固定的尿酸酶色谱柱即使在连续运行五个半月以上超过2000次后仍保留其原始活性的94%。