To evaluate the performance of intraocular lenses to treat cataract, an optomechanical eye model was developed. One of the most crucial components is the IOL holder, which should guarantee a physiological representation of the capsular bag and a stable position during measurement sequences. Individual holders are required due to the fact that every IOL has different geometric parameters. A method which allows obtaining the correct dimensions for the holder of a special IOL was developed and tested, by verifying the position of the IOL before and after a measurement sequence. Results of telecentric measurements and MTF measurements show that the IOL position does not change during the displacement sequence induced by the stepper motors of the eye model.
机译:为了评估人工晶状体治疗白内障的性能,开发了一种光机械眼模型。 IOL支架是最关键的组件之一,该支架应保证囊袋的生理学表现和在测量过程中的稳定位置。由于每个IOL都有不同的几何参数,因此需要单独的持有者。通过验证测量顺序前后IOL的位置,开发并测试了一种方法,该方法可以为特殊IOL的支架获取正确的尺寸。远心测量和MTF测量的结果表明,在由眼图模型的步进电机引起的位移序列中,IOL位置不变。

著录项

  • 来源
    《Biomedical imaging and sensing conference》|2017年|102511y.1-102511y.4|共4页
  • 会议地点 Yokohama(JP)
  • 作者单位

    Department of Biomedical, Health and Sports Engineering, University of Applied Sciences FH Technikum Wien, Hochstadtplatz 6,1200 Vienna, Austria;

    Department of Biomedical, Health and Sports Engineering, University of Applied Sciences FH Technikum Wien, Hochstadtplatz 6,1200 Vienna, Austria;

    Department of Biomedical, Health and Sports Engineering, University of Applied Sciences FH Technikum Wien, Hochstadtplatz 6,1200 Vienna, Austria;

    Department of Biomedical, Health and Sports Engineering, University of Applied Sciences FH Technikum Wien, Hochstadtplatz 6,1200 Vienna, Austria;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    eye model; intraocular lens; lens holder; IOL stability; optical measurement;