The objective of this research was to design and construct the control expiratory time adjustment display knob for the Bird Mark 7 ventilator. This research has adopted the principle of fluid mechanics and the principle of electronics. The designed and constructed control expiratory time adjustment display knob for the Bird Mark 7 ventilator was composed of 4 main parts: (1) the input part consists of pressure transducer, (2) the signal conditioner part composes of differential amplifier and voltage comparator, (3) the processing part consisting of MCS-51 P89V51RD2 microcontroller with C — language program and (4) the display part includes a liquid crystal display, which shows the duration of one cycle of breathing and the breathing rate. The results of functional testing compare with ventilator calibrator FLUKE VT — 30 found that the error of the average duration of a breathing cycle is equal to 1.36 percent and the average percentage error of the respiratory rate was 2 percent.
机译:这项研究的目的是设计和构造Bird Mark 7呼吸机的控制呼气时间调整显示旋钮。本研究采用了流体力学原理和电子学原理。 Bird Mark 7呼吸机的设计和构造的控制呼气时间调节显示旋钮由四个主要部分组成:(1)输入部分由压力传感器组成;(2)信号调节器部分由差分放大器和电压比较器组成;( 3)处理部分由带有C语言程序的MCS-51 P89V51RD2微控制器组成,并且(4)显示部分包括液晶显示器,该液晶显示器显示一个呼吸周期的持续时间和呼吸频率。功能测试的结果与呼吸机校准器FLUKE VT-30进行了比较,发现呼吸周期平均持续时间的误差等于1.36%,呼吸频率的平均百分比误差是2%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号