首页> 外文会议>60th Annual analysis division symposium 2015 >CLAUS SULFUR RECOVERY TAIL GAS APPLYING 100 MILLION HOURS OF OPERATIONAL TIME TO THE NEXT GENERATION ANALYZER
【24h】

CLAUS SULFUR RECOVERY TAIL GAS APPLYING 100 MILLION HOURS OF OPERATIONAL TIME TO THE NEXT GENERATION ANALYZER

机译:克劳斯硫磺回收尾气将1亿小时的运行时间应用于下一代分析仪

获取原文
获取原文并翻译 | 示例

摘要

The measurement of H_2S/SO_2 in Claus sulfur recovery unit (SRU) tail gas has been adequately addressed by UV spectroscopy for over 40 years. Reliability of the analytical principle was established in the first generation of analyzers and in the second generation sample handling was improved to the point where automatic control if not universal is at least considered the norm. With a deep understanding of the process and a population of 1,100 second generation analyzers it was possible in the third generation to address failure modes external to the analyzer. Reliability is not limited to the analyzer / sample system; it extends to the process and contains elements of health and safety. Analyzer professionals are compelled to look beyond what we consider our "deliverables", to address abnormal operations and bad piping design. The paper combines extensive feedback from analyzer professionals and a survey of sulfur recovery operators to address the external failure modes. Looking back on 100 million hours of operational time and one year of field testing the third generation analyzer the paper discusses reliability as viewed by the real end use customer; Operations.
机译:紫外光谱已充分解决了克劳斯硫磺回收装置(SRU)尾气中H_2S / SO_2的测量问题,已有40多年的历史了。在第一代分析仪中建立了分析原理的可靠性,在第二代分析仪中,样品处理得到了改进,以至于至少将自动控制视为通用。通过对过程的深刻理解和1100台第二代分析仪的使用,第三代分析仪可以解决分析仪外部的故障模式。可靠性不仅限于分析仪/样品系统;它扩展到整个过程,并包含健康和安全要素。分析仪专业人员必须超越我们认为的“可交付成果”范围,以解决异常操作和不良管道设计的问题。本文结合了分析仪专业人员的广泛反馈和对硫磺回收操作员的调查,以解决外部故障模式。回顾第三代分析仪的1亿小时的运行时间和一年的现场测试,本文讨论了最终用户所看到的可靠性。操作。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有